Verification of field placement for rotational IMRT
M. Saiful Huq, Greg Bednarz, Murshed Hossain, James Galvin, Walter J. Curran, Jr.
Radiation Oncology, Thomas Jefferson University
Philadelphia, PA Techniques for guaranteeing that intensity patterns are properly
positioned relative to a patients anatomy have not been developed for rotational
IMRT. This paper points out the importance of at least assuring that the "slit"
opening covers the treatment volume. Using the NOMOS MIMiC collimator, it is possible, and
sometimes necessary, to position the treatment unit isocenter so that the collimator
opening does not provide full target coverage as the gantry rotates. This report
demonstrates that missing the target compromises the dose distribution by forcing
normalization to relatively low isodose lines and increasing the dose to critical healthy
structures. A simple check of the NOMOS output can detect potential problems. Caution is
advised when the beam elements at the field edge are used in the modulation process. As a
more complete check, a double-exposure film technique using the standard jaws is
recommended to document field coverage. The technique uses both an anterior and one
lateral film to show the extent of the collimator opening. The field size for these films
is set to agree with the length of the beam slit (approximately 20 cm) in one direction
and the slit width in the other direction. A double-exposure over-flash is used to show
this slit relative to the patients anatomy. Simulation of these fields is also
recommended. Since CT scans must be available for any IMRT treatment, the use of
CT-simulation simplifies this part of the process.
Verification of field placement for rotational IMRT
The NOMOS MIMiC intensity modulating device is a tertiary collimator and clearance can
be a problem. Installation on a Philips SL 75/5 allows only 29 cm from the front face of
the collimator to the treatment unit isocenter. This distance sometimes forces placement
of the mechanical isocenter near the surface of the tabletop to avoid couch collision as
the gantry rotates. However, the number of leaves limits the length of the slit opening
for this device. For the SL 75/5 unit, this length is 18 cm. If a lesion plus margin
extends more than 9 cm from the position of the isocenter, it is not be possible to
deliver radiation to some parts of the target for one or more portions of the arc. This
compromises the optimization and results in reduced dose homogeneity relative to the
situation where complete coverage is possible. There are no warnings within the planning
software or its documentation to caution the dosimetrist that a sub-optimal plan has been
generated. Since verification of IMRT treatments is not standard, this problem can go
undetected. Using film dosimetry to verify the calculated treatment plan against
measurement will not demonstrate the problem because the two poor plans will agree. For
this reason, simple checks are suggested here to detect situations where complete coverage
by the collimator is not achieved.

The problem is illustrated for a patient where it was not possible to position the
isocenter so that the collimator opening covered the anterior portion of the lesion when
the gantry was in the lateral position. Dose-volume histograms (DVH) for the plan where
collision was a problem are compared to DVHs obtained when the isocenter was placed
near the center of the target. Due to the possibility of collision, the second plan is
achievable with our equipment. The figure above compares the two treatment plans. Both
plans cover 90% of the GTV with the boost dose of 24Gy. Notice that the dose inhomogeneity
for the plan in the upper part of the figure is much greater than the inhomogeneity for
the plan where the field opening fully covers the target. Also notice that the dose for
all critical structures is increased for the plan where the field misses the target.
A number of checks can be performed to guarantee target coverage. A simple check is to
observe the intensity patterns printed by the NOMOS CORVUS software. If dose delivery
extends to the very edge of the collimator opening, some shielding of the target should be
suspected. Port filming of the available field opening is also recommended. The filming
technique developed here uses a double exposure without the NOMOS collimator in place. The
first exposure shows the patients anatomy with extensive margin, and the second
reduces the collimator size to correspond to the MIMiC opening. A single lateral field
film and an anterior film are used to demonstrate acceptable coverage. If the printed
intensity patterns show the possibility of missing the target at angles other than 0 and
90 degrees, additional angles should be included in the filming. The figure below shows an
example of a lateral port film. Notice that the field opening does not extend to the
anterior portion of the patient where there is disease. At our institution, the treatment
couch was modified to allow shifting of the isocenter to a greater distance above the
tabletop to provide a larger range for positioning the irradiated volume.
It should be pointed out that some clipping of the target may be acceptable in that
dose delivered form other directions can possibly make up for projections where the volume
is not irradiated. However, it is always prudent to generate a plan with full coverage to
determine if the optimization is seriously compromised.

Abstracts Submitted to 1999 Annual AAPM Meeting | Medical Physics Home |