An Evaluation of the Recommendations of the
AAPMapm TG51 Protocol: Photon Beam Calibration And Comparison With the TG21 Protocol
M. Saiful Huq*
Radiation Oncology
Jefferson Medical College of Thomas Jefferson University
Philadelphia, PATask Group 51 of the AAPM Radiation Therapy Committee has recently
developed a new protocol for the calibration of high-energy photon and electron beams
based on standards of absorbed dose to water. The recommended dosimetry procedures in this
protocol are based on the use of an ionization chamber calibrated in terms of absorbed
dose to water in a standards laboratorys 60Co beam. This is different
from the recommendations given in the TG21 protocol, which are based upon an exposure
calibration factor of an ionization chamber in a 60Co beam. The purpose of this
work is to compare the recommendations of the two protocols using two widely used Farmer
type ionization chambers: PTW N30001 (PMMA wall) and NEL 2571 (Graphite wall). This was
accomplished by determining absorbed-dose-to-water with both chambers irradiated in a
water phantom using 6 and 25 MV photon beams. Ratios of TG51 to TG21 doses were found to
be 1.006 and 1.007 at 6 and 25 MV respectively when the PTW chamber is used; 1.010 and
1.011 when the NEL chamber is used. When the recommendations of TG51 are followed the dose
determined by the NEL chamber is found to be higher than that of the PTW chamber by 1.6%
at 6 MV and 1.8% at 25 MV. The corresponding results for the TG21 protocol are 1.2% at 6
MV and 1.4% at 25 MV respectively.
An Evaluation of the Recommendations Of The AAPM TG51 Protocol:
Photon Beam Calibration And Comparison With the TG21 Protocol
Comparison of Absorbed Doses-to-Water Determined by Following the
Recommendations of the TG51 and TG21 Protocols.